Increased image quality and reduced radiation dose and contrast media volume: a holistic approach to intracranial CTA
Clinical Radiology, Volume 72, No. 9, Year 2017
Notification
URL copied to clipboard!
Aim To investigate the dose–length product (DLP) during intracranial computed tomography angiography (CTA) using a patient-specific contrast formula. Materials and methods Intracranial CTA was performed on 120 patients using 64-channel CT. Patients were subjected in equal numbers to one of two acquisitions/contrast medium protocols. Protocol A, consisted of 80 ml contrast medium and protocol B, involved a novel contrast medium formula. In each protocol, contrast medium and saline were injected at a flow rate of 4.5 ml/s. The DLP and contrast volume (CV) were measured between each protocol and the data obtained were compared using two-tailed independent t-test. Results Mean arterial vessel attenuation was up to 56% (p<0.01) higher using protocol B compared with A. In the venous system, the mean vessel attenuation was significantly lower in protocol B than A with a maximum reduction of 93% (p<0.001). The mean CV was significantly lower in protocol B (53±10 ml) compared to A (80±1 ml, p<0.001). The scan time was equal in each protocol (B, 4.22±1.2 seconds; A, 4.01±1.3 seconds). A significant reduction in mean DLP was demonstrated in protocol B (3.99±0.22 mSv) compared to A (4.74±0.22 mSv; p=0.02). Conclusion A significant reduction in CV and DLP during intracranial CTA can be achieved when employing a patient-specific contrast medium formula.