Publication Details

AFRICAN RESEARCH NEXUS

SHINING A SPOTLIGHT ON AFRICAN RESEARCH

medicine

Variability of anti-human transglutaminase testing in celiac disease across Mediterranean countries

World Journal of Gastroenterology, Volume 23, No. 24, Year 2017

AIM To verify the precision and accuracy of transglutaminase antibodies (TGA) assays across Mediterranean countries. METHODS This study involved 8 referral centres for celiac disease (CD) in 7 Mediterranean countries. A central laboratory prepared 8 kits of 7 blinded and randomized serum samples, with a titrated amount of Human TGA IgA. Each sample was analysed three times on three different days, with each centre running a total of 21 tests. The results were included in a blindly coded report form, which was sent to the coordinator centre. The coordinator estimated the mean coefficient of Variation (CoVar = σ/μ), the mean accuracy (Accur = Vobserved -Vreal) and the mean percent variation (Var% = [(Vobserved -Vreal)/Vreal] × 100). RESULTS The analysis showed that 79.17% of the mean variation fell between -25% and +25% of the expected value, with the accuracy and precision progressively increasing with higher titres of TGA. From values 1.25 times greater than the normal cut-off, the measurements were highly reliable. CONCLUSION TGA estimation is a crucial step for the diagnosis of CD; given its accuracy and precision, clinicians could be confident in establishing a diagnosis.
Statistics
Citations: 12
Authors: 12
Affiliations: 10
Identifiers