Fiberoptic light source-induced surgical fires - The contribution of forced-air warming blankets
Acta Anaesthesiologica Scandinavica, Volume 50, No. 4, Year 2006
Notification
URL copied to clipboard!
Description
Background: Fiberoptic light sources have been identified as a fire ignition mechanism in the operating room. This study attempted to determine whether a forced-air warming blanket (FAWB) could affect the ignition or spread of fire caused by a fiberoptic light source. Methods: We exposed surgical drapes to a fiberoptic light source at close range. The results were categorized according to time to first smoke and damage resulting at 1 min. Data were analyzed using the Mann-Whitney rank-sum test. Results: The sums of the rank values for the components of the drape indicated that there was a greater than 96.8-99.2% chance that the FAWB accelerated the time to first smoke. The FAWB appeared to protect the patient gown from damage during all trials. The presence of an FAWB under a surgical drape accelerated the time to first smoke when exposed to unprotected fiberoptic light sources, yet prevented damage to the underlying patient gown. In an actual surgical setting, it is likely that the FAWB would offer some protection to the patient's skin directly below the surgical drape. Conclusion: It is likely that the FAWB can offer some protection to the patient should an unprotected fiberoptic light source cause a fire. © 2006 Acta Anaesthesiol Scand.
Authors & Co-Authors
Williams, David Michael J.
United States, San Jose
Santa Clara Valley Medical Center
Patterson, Andrew J.
United States, Palo Alto
Stanford University
Brock-Utne, John Gerhard
United States, Palo Alto
Stanford University
Statistics
Citations: 17
Authors: 3
Affiliations: 3
Identifiers
Doi:
10.1111/j.1399-6576.2006.00975.x
ISSN:
13996576
Research Areas
Environmental
Health System And Policy