Patients exposure and imaging quality in chest radiographs: A critical evaluation; [Exposição de pacientes e qualidade da imagem em radiografias de tórax: Uma avaliação crítica]

Radiologia Brasileira, Volume 40, No. 2, Year 2007

OBJECTIVE: Entrance skin dose, effective dose, and imaging quality in chest radiographs of adult patients have been evaluated. MATERIALS AND METHODS: The study has been developed in eight institutions - seven public hospitals (two of them philanthropic institutions) and one private - in the cities of Agra dos Reis, Cabo Frio, Campos dos Goytacazes, Itaperuna, Niterói, Recife and Rio de Janeiro. Entrance skin dose and effective dose have been evaluated in 735 chest radiographs obtained in posteroanterior/anteroposterior and lateral projections. As regards imaging criteria, 44 radiographs have been evaluated. RESULTS: Variations of up to nine times in entrance skin dose, and six times in effective dose have been detected for a same type of projection. Also, significant discrepancies have been found in values resulting from radiographic techniques employed. Besides, imaging quality has not been good since the rate of compliance with imaging criteria was only 55%. CONCLUSION: There is a pressing need for improvement/standardization of procedures in conventional radiology; this can be achieved by implementing a quality control and assurance program in the department of radiology, including training of technicians, x-ray equipment calibration, and sensitometric control of films processors.