Comparative dose levels between CT-scanner and slot-scanning device (EOS system) in pregnant women pelvimetry

Physica Medica, Volume 33, Year 2017

Purpose To estimate fetal absorbed doses for pregnant women pelvimetry, a comparative study between EOS imaging system and low-dose spiral CT-scanner was carried out. For this purpose three different studies were investigated: in vivo, in vitro and Monte Carlo calculations. Methods In vivo dosimetry was performed, using OSL NanoDot dosimeters, to determine the dose to the skin of twenty pregnant women. In vitro studies were established by using a cubic phantom of water, in order to estimate the out of field doses. In the latter study, OSLDs were placed at depths corresponding to the lowest, average and highest position of the uterus. Monte Carlo calculations of effective doses to high radio-sensitive organs were established, using PCXMC and CTExpo software suites for EOS imaging system and CT-scanner, respectively. Results The EOS imaging system reduces radiation exposure 4 to 8 times compared to the CT-scanner. The entrance skin doses were 74% (p-values <0.01) higher with the CT-scanner than with the EOS system. In the out of field region, the measured doses of the EOS system were reduced by 80% (p-values <0.02). Monte Carlo calculations confirmed that effective doses to organs are less accentuated for EOS than for CT pelvimetry. Conclusions The EOS system is less irradiating than the CT exam. The out-of-field dose which is significant, is lower in the EOS than in the CT-scanner and could be reduced even further by optimizing the time used for image acquisition.
Statistics
Citations: 15
Authors: 6
Affiliations: 5
Identifiers
Research Areas
Cancer
Environmental
Participants Gender
Female